        YLN-PCSE-2002

        YLN-PCSE-2002 Five-Parameter Water Quality Automatic Online Monitor

        Communication Device:1 Sampling cycle: time interval (arbitrarily adjustable from 1 to 9999min) and hourly measurement mode.2 Calibration cycle: adjustable at any interval any time from 1 to 99 days.3 Maintenance cycle: 1 time/month.4 Display: 10-inch…

        Order Online
        • Product Principle
        • Product Features
        • Technical Parameters

        Communication Device:

        1 Sampling cycle: time interval (arbitrarily adjustable from 1 to 9999min) and hourly measurement mode.

        2 Calibration cycle: adjustable at any interval any time from 1 to 99 days.

        3 Maintenance cycle: 1 time/month.

        4 Display: 10-inch color screen, 1024*600, 16.77 million colors

        5 Output: RS-232, 4-20mA (optional).

        6 Communication protocol: Modbus RTU, new national standard protocol, dynamic control protocol, others

        7 Environment requirements: indoor with adjustable temperature, recommended temperature +5~28℃; humidity ≤90% (no condensation).

        8 Power supply: AC220±10% V, 50±10% Hz, 5A.

         

        Communication Device:

        1 Sampling cycle: time interval (arbitrarily adjustable from 1 to 9999min) and hourly measurement mode.

        2 Calibration cycle: adjustable at any interval any time from 1 to 99 days.

        3 Maintenance cycle: 1 time/month.

        4 Display: 10-inch color screen, 1024*600, 16.77 million colors

        5 Output: RS-232, 4-20mA (optional).

        6 Communication protocol: Modbus RTU, new national standard protocol, dynamic control protocol, others

        7 Environment requirements: indoor with adjustable temperature, recommended temperature +5~28℃; humidity ≤90% (no condensation).

        8 Power supply: AC220±10% V, 50±10% Hz, 5A.

        Water Temperature Technical Parameters  

        1. Measuring principle: temperature sensor

        2. Measuring range: -5~60℃

        3. Accuracy: ±0.5℃

        4. Resolution: 0.1℃

        5. MTBF: ≥720h/time

        6. Output signal: 4-20mA, RS-485/232

         

        pH Technical Parameters

        1. Test principle: glass electrode method

        2. Measuring range: 0.00~14.00 pH

        3. Repeatability: within ±0.01pH

        4. Resolution: 0.01 pH

        5. Temperature compensation: automatic temperature compensation function

        6. Function: glass electrode damage self-checking function

        7. MTBF: ≥720h/time

        8. Comparison test with actual water sample: within ±0.1pH

        9. Output signal: 4-20mA, RS-485/232

         

        Dissolved Oxygen Technical Parameters

        1. Measurement method: membrane electrode method or chemiluminescence method

        2. Measuring range: 0.00~20.00mg/L

        3. Repeatability error: ±0.3 mg/L

        4. Resolution: 0.05mg/L

        5. Response time: 90% guarantee rate≤60S

        6. Temperature compensation: automatic temperature compensation function

        7. Drift: ≤0.3mg/L

        8. Function: with electrode self-check function. Membrane electrode method can detect membrane damage and electrolyte leakage.

        9. Comparison test with actual water sample: ±0.3 mg/L

        10. MTBF: ≥720h/time

        11. Output signal: 4-20mA, RS-485/232

         

        Electrical Conductivity Technical Parameters

        1. Test principle: electrode method

        2. Measuring range: 0~20mS/cm, measuring range can be switched automatically

        3. Repeatability error: ±1%

        4. Resolution: ≤0.1μS/cm

        5. Temperature compensation: automatic temperature compensation function

        6. MTBF: ≥720h/time

        7. Output signal: 4-20mA, RS-485/232

        8. Comparison test with actual water sample: ±10%

        9. Drift: ±1%

         

        Turbidity Technical Parameters

        1. Test principle: using 90 degree scattering method

        2. Measuring range: 0-4000NTU, the range can be automatically switched

        3. Repeatability error: ±5%

        4. Resolution: 0.01NTU

        5. Function: with self-checking function and automatic cleaning function

        6. MTBF: ≥720h/time

        7. Output signal: 4-20mA, RS-485/232

        8. Comparison test with the actual water sample, the relative test error≤10%